9-11
9 CHECK AND ADJUSTMENT
9.5 Showing Constant List and Switch ON/OFF
Compensation Systematic Error of Instrument
[Example setting: Switch OFF the compensation]
Operating procedure Operation Display
1
Press [F5] key from the main menu.
[F5]
2
Press [F2] key .
Correction values are displayed.
[F2]
3
Press [F6](OFF) key.
[F6]
4
Press [F1](EXIT) key.
The display returns to main menu.
[F1]
Adjustment
F1 V0/Axis (Measurement)
F2 V0/Axis (Constant list)
F3 Date Time
F4 Instrument constant
Vco: -1°57'12"
Hco: -0°00'20"
HAx: -0°00'20"
EXIT ON OFF
Vco: -1°57'12"
Hco:
HAx:
EXIT ON OFF
Vue de la page 109